4.10. Factory Restore

The ALx Series MagnaLOAD DC electronic load contains EEPROM (electrical erasable programmable read-only memory) for retaining settings after loss of AC power. The memory contains the set points, control mode, calibration gains, offsets, product serial numbers, and more. Most the memory positions are visible using MagnaWEB under the “EEPROM Editor” side menu.

Factory Restore overwrites existing EEPROM settings with defaults values so the MagnaLOAD electronic load can return to a known operating state. This should be performed if user experiences unexpected behavior, due to unknown or forgotten configurations. Magna-Power Electronics may request users perform factory restore, as a starting point in most support cases. The memory positions affected by factory restore are listed in the table below.

Factory restore values (defaults) can be stored in Flash or EEPROM memory, as determined by Magna-Power Electronics. The Flash values are used for product-wide settings, such as control mode and slew rates, and mostly affect the products operating state. EEPROM values are reserved for customer and unit-specific settings, such as gains, offsets, and MagnaLINK™ address information, which are set during factory testing.

Product can be restored to a known state, by modifying the EEPROM in bulk, using either Soft Restore or Hard Restore. Soft Restore only effects those EEPROM positions whose default values are stored in Flash, as shown in the table below. Hard Restore overwrites calibrations values, feedback gains, i.e., settings saved in EEPROM at the factory. In most cases, Soft Restore should be used, and Hard Restore used as a last resort, since it overwrites any in-house calibrations with or original (older) factory calibrations.

Factory restore can be initiated using the front panel by navigating to System Settings - Factory Restore in the menu system. After the restore option has been selected, MagnaLOAD electronic load will reboot and start up with the new settings.

Factory Restore Values

Name

Memory Type

Restore Value

Description

setPointReg0

flash

0

Current setpoint (normalized)

setPointReg1

flash

0

Voltage setpoint (normalized)

setPointReg2

flash

0

Power setpoint (normalized)

setPointReg3

flash

0

Resistance setpoint (normalized)

setPointReg4

flash

0

Rheostat setpoint (normalized)

tripOverVoltage

flash

1.1

Over Voltage Trip (normalized)

tripUnderVoltage

flash

0

Under Voltage Trip (normalized)

tripDelayVoltage

flash

0

Delay Voltage Trip (milliseconds)

tripOverCurrent

flash

1.1

Over Current Trip (normalized)

tripUnderCurrent

flash

0

Under Current Trip (normalized)

tripDelayCurrent

flash

0

Current Trip Delay (milliseconds)

tripOverPower

flash

1.1

Over Power Trip (normalized)

tripUnderPower

flash

0

Under Power Trip (normalized)

tripDelayPower

flash

0

Power Trip Delay (milliseconds)

tripOverResistance

flash

0

Over Resistance Trip (normalized)

tripUnderResistance

flash

0

Under Resistance Trip (normalized)

tripDelayResistance

flash

0

Resistance Trip Delay (milliseconds)

setpointSelect

flash

1

Last setpoint selected by user

controlMode

flash

1

Regulation control mode

rampRiseVoltage

eeprom

factory determined

Voltage setpoint rise rate (1/s)

rampRiseCurrent

eeprom

factory determined

Current setpoint rise rate (1/s)

rampRiseResistance

eeprom

factory determined

Resistance setpoint rise rate (1/s)

rampRisePower

eeprom

factory determined

Power setpoint rise rate (1/s)

rampFallVoltage

eeprom

factory determined

Voltage setpoint rise fall (1/s)

rampFallCurrent

eeprom

factory determined

Current setpoint rise fall (1/s)

rampFallResistance

eeprom

factory determined

Resistance setpoint rise fall (1/s)

rampFallPower

eeprom

factory determined

Power setpoint rise fall (1/s)

range

flash

0

Power range, controls behavior of bypassing resistor modules (LowPower:0, HighPower:1)

idxHighSpeedAI

flash

0

Selection index that chooses which setpoint, high-speed analog-in signal is applied to.

functionType

flash

0

Function Type Selection. See EnumFunctionType for more details

funcSineAmpl

flash

0.1

Sine Function Amplitude (normalized), applies to Current setpoint.

funcSineOffset

flash

0.5

Sine Function Offset (normalized), applies to Current setpoint.

funcSinePrd

flash

10

Sine Function Period in milliseconds

funcSqLoLevel

flash

0.1

Square Function Low Level (normalized), applies to Current setpoint.

funcSqHiLevel

flash

0.5

Square Function High Level (normalized), applies to Current setpoint.

funcSqLoPrd

flash

10

Square Function Low Period in milliseconds

funcSqHiPrd

flash

10

Square Function High Period in milliseconds

funcStepLoLevel

flash

0.1

Step Function Low Level (normalized), applies to Current setpoint.

funcStepHiLevel

flash

0.5

Step Function High Level (normalized), applies to Current setpoint.

funcRampLoLevel

flash

0.1

Ramp Function Low Level (normalized), applies to Current setpoint.

funcRampHiLevel

flash

0.5

Ramp Function High Level (normalized), applies to Current setpoint.

funcRampRisePrd

flash

10

Ramp Function Rise Period in milliseconds

funcRampFallPrd

flash

10

Ramp Function Fall Period in milliseconds

setSource

flash

0

Select Setpoint Source from Local, Function or External Ctrl

sampleMode

eeprom

factory determined

Change sampling mode on communication bus. 0 returns filtered samples, 1 return unfiltered values

senseMode

eeprom

factory determined

Voltage terminals measurements sourced from local sense (0) or remote sense (1) or leadless sense(2)

lockMode

eeprom

factory determined

Behavior of unit when lock button is pressed

name

eeprom

factory determined

User assigned name for unit

tapSetting

eeprom

factory determined

Amount of digital filtering for voltage and current. Restricted between 0 (no filtering-fast) to 30 (full filtering-slow)

ctrlModeAuxPwr

eeprom

factory determined

Control modes for auxiliary power supply.

protocol

eeprom

factory determined

Select protocol for communications between product and computer.

sysMode

eeprom

factory determined

System Mode, Configuration in which system of units are connected (standalone:0, parallel:1, series:2)

productModelNum

eeprom

factory determined

Product model number (e.g., ARx7.5-1000-15)

productType

eeprom

factory determined

Product type (e.g., ARx, ALx, WRx)

numFans

eeprom

factory determined

Number of fans installed in the unit

frontPanelAddr

eeprom

factory determined

Front panel target address (MagnaLINK)

frontPanelSeriesNum

eeprom

factory determined

Front panel series number

frontPanelSerialNum

eeprom

factory determined

Front panel serial number, Zero if C-Panel

ratedVoltTarget

eeprom

factory determined

Maximum voltage (V) target is specified to operate at

ratedCurrTarget

eeprom

factory determined

Maximum current (A) target is specified to operate at

ratedResTarget

eeprom

factory determined

Maximum resistance (ohm) target is specified to operate at

ratedPwrTarget

eeprom

factory determined

Maximum power (W) target is specified to operate at

ratedInputVoltTarget

eeprom

factory determined

Maximum line-to-line input voltage (Vrms) product is specified to operate at

ratedInputCurrTarget

eeprom

factory determined

Maximum input current (Arms) product is specified to operate at

ratedInputVoltTrip

eeprom

factory determined

Input voltage phase loss trip constant

ratedRamp

eeprom

factory determined

[Maximum Voltage Slew Rate; Maximum Current Slew Rate; Maximum Power Slew Rate; Maximum Resistance Slew Rate]

ratedVTermSensor

eeprom

factory determined

Rating for the terminal voltage sensor (V)

ratedITermSensor

eeprom

factory determined

Rating for the terminal current sensor (A)

ratedVDcLinkSensor

eeprom

factory determined

Rating for the dc link voltage sensor (V)

ratedILSensor

eeprom

factory determined

Rating for the phase current sensors (A)

ratedCurrChopper

eeprom

factory determined

Rating for the phase current(A) at rated power

nChopperPhase

eeprom

factory determined

Number of chopper modules

nLinePhase

eeprom

factory determined

Number AC input phases from the utility

chopperPrd

eeprom

factory determined

Chopper Frequency Period in terms of number of Clock Cycles, defaults to 9.6kHz

chopperClkDiv

eeprom

factory determined

Chopper Frequency Clock Scaler Divisor

inverterPrd

eeprom

factory determined

Inverter Frequency Period in terms of number of Clock Cycles, defaults to 600Hz

inverterClkDiv

eeprom

factory determined

Inverter Frequency Clock Scaler Divisor

gainP1AutoX

eeprom

factory determined

Autocrossover compensator proportional gain1

gainI1AutoX

eeprom

factory determined

Autocrossover compensator integral gain1

gainP2AutoX

eeprom

factory determined

Autocrossover compensator proportional gain2

gainI2AutoX

eeprom

factory determined

Autocrossover compensator integral gain2

gainPModeChopper

eeprom

factory determined

Chopper compensator proportional gain

gainIModeChopper

eeprom

factory determined

Chopper compensator integral gain

gainDModeChopper

eeprom

factory determined

Chopper compensator derivative gain

gainPModeIPV

eeprom

factory determined

IPV compensator proportional gain

gainIModeIPV

eeprom

factory determined

IPV compensator integral gain

gainPModeVPI

eeprom

factory determined

VPI compensator proportional gain

gainIModeVPI

eeprom

factory determined

VPI compensator integral gain

gainPModeRPIV

eeprom

factory determined

RPIV compensator proportional gain

gainIModeRPIV

eeprom

factory determined

RPIV compensator integral gain

gainPModePIV

eeprom

factory determined

PIV compensator proportional gain

gainIModePIV

eeprom

factory determined

PIV compensator integral gain

gainPModeShuntReg

eeprom

factory determined

Shunt regulator compensator proportional gain1

gainIModeShuntReg

eeprom

factory determined

Shunt regulator compensator integral gain1

gainBModeShuntReg

eeprom

factory determined

Shunt regulator compensator band

gainCalibrate

eeprom

factory determined

[Voltage Gain; Current Gain; Remote Sense Voltage Gain; Phase Current Gain]

gainCalibrateCtrl

eeprom

factory determined

[Set-point PWM Gain; Capture PWM Gain; available; available; available]

offset0

eeprom

factory determined

Protection sensor offset

offset1

eeprom

factory determined

Terminal voltage sensor offset

offset2

eeprom

factory determined

Terminal remote voltage sensor offset

offset3

eeprom

factory determined

Terminal current sensor offset

offset4

eeprom

factory determined

Phase current 1 sensor offset

offset5

eeprom

factory determined

Phase current 2 sensor offset

offset6

eeprom

factory determined

Phase current 3 sensor offset

offset7

eeprom

factory determined

Linear module temperature sensor offset

offset8

eeprom

factory determined

Resistor module 1 temperature sensor offset

offset9

eeprom

factory determined

Resistor module 2 temperature sensor offset

offset10

eeprom

factory determined

External analog set point

offset11

eeprom

factory determined

Internal ADC offset Main Control

offset12

eeprom

factory determined

Internal ADC offset Gate Drive

offset13

eeprom

factory determined

Internal ADC offset Auxiliary Power

offset14

eeprom

factory determined

Control PWM offset

availAddrLocal

eeprom

factory determined

Address of devices on local network (port A)

commOption

eeprom

factory determined

Communication daughter board type

ratedVoltProduct

eeprom

factory determined

Maximum voltage (V) product is specified to operate at

ratedCurrProduct

eeprom

factory determined

Maximum current (A) product is specified to operate at

ratedResProduct

eeprom

factory determined

Maximum resistance (Ohm) product is specified to operate at

ratedPowerProduct

eeprom

factory determined

Maximum power (W) product is specified to operate at

ratedVoltSystem

eeprom

factory determined

Maximum voltage (V) system is specified to operate at

ratedCurrSystem

eeprom

factory determined

Maximum current (A) system is specified to operate at

ratedResSystem

eeprom

factory determined

Maximum resistance (Ohm) system is specified to operate at

ratedPowerSystem

eeprom

factory determined

Maximum power (W) System is specified to operate at

magnaRouter

eeprom

factory determined

Enable MagnaLINK Interface Device

enPortSerialCheck

eeprom

factory determined

Set to enable serialized messages on any port

factoryValueSet

eeprom

factory determined

Set to EEPROM to factory calibration mode (disk zero)

factoryRestoreMode

eeprom

factory determined

Restore factory defaults to current disk

optionRegister

eeprom

factory determined

Integrated Options available on a product

deratedVoltProdFactor

eeprom

factory determined

Derated voltage product is limited to operate at

deratedCurrProdFactor

eeprom

factory determined

Derated current product is limited to operate at

deratedResProdFactor

eeprom

factory determined

Derated resistance product is limited to operate at

deratedPowerProdFactor

eeprom

factory determined

Derated power product is limited to operate at

productConfig

eeprom

factory determined

Product Configuration (e.g., A1, B1, C1)

featureRegister

eeprom

factory determined

Features available on a product

statusRegister0Mask

eeprom

factory determined

Mask status fault bits. See EnumStatusRegister for more information.

statusRegister1Mask

eeprom

factory determined

Mask status fault bits. See EnumStatusRegister for more information.

funcPeriodMin

eeprom

factory determined

Minimum Function Period in milliseconds

funcPeriodMax

eeprom

factory determined

Maximum Function Period in milliseconds

sequenceId

eeprom

factory determined

Sequence Identification

cmdStartAddr

eeprom

factory determined

Storage location of sequence commands

cmdLength

eeprom

factory determined

Length of commands programmed

paramStartAddr

eeprom

factory determined

Storage location of sequence parameters

paramLength

eeprom

factory determined

Length of parameters programmed